Mississippi Polymer InstituteWestern Carolina UniversityDynamic Systems Inc.
Mississippi Polymer Institute
Research Report
ACS Report
Graphs
AccuCure Setup
Gel Setup



Measurement of the Extent of Reaction During Polymerization Using DSC, Gel Time and a Newly Discovered Electrical Phenomena

Tracy Wyatt and Tison Wyatt
Tison Technologies, LLC
PO Box 1268

Skyland, N.C. 28776-1268

Phone: 828-684-0050
E-mail: info@tison.com

Kenneth T. Posey and Richard D. Sudduth
Mississippi Polymer Institute

University of Southern Mississippi

Box 10003

Hattiesburg, Ms. 39406

Phone 601-266-4607

As announced at :

American Chemical Society
49th Southern Regional Meeting

Roanoke, Virginia

October 19-22, 1997

The Discovery

Most chemical reactions produce a voltage that can be detected passively utilizing a patented electrode.

  • This technique has been called AccuCure™

What is AccuCureTM

  • Electrodes must be adequately shielded and grounded from spurious voltages.

AccuCure™, using techniques which are patented, measures voltage produced by a chemical reaction, i.e. polymerization, etc.

  • Voltage apparently indicative of the state of electrons being transferred during a reaction
  • Domains of Cathodic and Anodic regions
  • need to be large enough to detect measurement
  • Domains must not be too random (good for Thermosets).

Materials Evaluated

Epoxy (Polycondensation Polymerization)

- EPON Resin 828

- EPI-CURE 3140 Polyamide Curing Agent

Urethane (Polycondensation Polymerization)

- Smooth-On Crystal Clear 200 Part A

- Smooth-On Crystal Clear 200 Part B

Unsaturated Polyester (Free Radical Polymerization)

- Bondo (Unsaturated Polyester/Styrene Monomer)

- Benzoyl Peroxide

Formulated Reaction Rate

Considerations

Gel-Times approximately 8 hrs. at room temperature

Long gel time has the advantage of minimizing weighting and mixing time errors

Gel-Time approximately ten minutes at 150 F or 180 F depending on reacting medium.

Short gel times are advantageous for study simply because the rate of electron transfer is increased

Experimental Techniques

  • AccuCureTM Probe Measurement

(Isothermal)

- AccuCure™ Process

  • DSC Measurements (Isothermal)

- Perkin-Elmer Model # DSC-7

  • Gel-Time Measurement (Isothermal)

- Sunshine Gel Checker 22-B

AccuCure™ Measurement Results

  • AccuCure measurement reproducible and consistent
  • AccuCure produced mirror voltage curves for urethane and epoxy
  • Did not get mirror voltage curve for unsaturated polyester polymerization
  • Maximum for polyester polymerization appeared to occurred at gel-time
  • Location of gel-time for epoxy and urethane consistent relative to AccuCure curve

Comparison of Reaction Types

· Condensation or step wise reactions evaluated
- Polyurethane reaction
- Epoxy reaction
· Condensation or step wise crosslinking characteristics.
- Early in the reaction monomer or low molecular weight units disappear rapidly.
>> Reaction becomes ordered quickly.
- Since each step in the polymerization proceeds by approximately the same rate, as reactive sites are used up the polymerization rate decreases

· Free radical reaction evaluated
- Unsaturated polyester reaction free radical crosslinking characteristics
- Chain polymerization reaction process is very complex with many reactive processes occurring simultaneously.
>> Initiation Stage
>> Propagation Stage
>> Termination Stage
- At any instance in reaction the mixture could contain monomer, high MW polymer, and growing chains

DSC Measurement Compared

to Identical Electrodes Measurement

  • DSC exothermic energy assumed to be one measure of rate at which electrons are being transferred during a reaction
  • Since the AccuCure process is also believed to involve the transfer of electrons some correlation with the DSC results were expected

Conclusions

  • AccuCure and DSC indicate a minimum of one corresponding peak for polyurethane and epoxy reactions
  • AccuCure TM and DSC did not indicate a corresponding peak for the polyester reaction
  • AccuCure appears to be able to distinguish a different reaction spectrum for each reaction type
  • Gel-Time is indicated differently by AccuCure for the three different reactions
  • AccuCure is a measurement tool with great potential to measure reaction progress
  • AccuCure shows remarkable consistency, even with mirror image results
  • Phenomena measured using the AccuCure process is not yet fully understood.

However, if AccuCure produces consistent, reproducible and usable results, a detailed understanding of the voltage phenomena is not required.

Potential Applications for

AccuCureTM

  • Unstirred chemical reactions after initial mixing

- Thermosets (Composites)
- Rubbers / Elastomers

  • Reaction identification or reaction spectrum
  • Measurement of critical reaction phases

- Reaction Onset
- Reaction Rate

- Gel Time
- Completion of Multiple Reaction Scheme

- Cure Cycle

Major Advantages of AccuCure's

Passive Feature

  • Process itself determines process completion
  • Data Collection can be easily achieved during actual production runs without disturbing finished product
  • Since AccuCure™ is a passive monitoring system, potentially dangerous and/or explosive reactions can be monitored with minimal safety concerns

Potential Production Advantages

of AccuCureTM

  • Reduce cycle time
  • Reduction of lost production due to incorrect processing
  • Ability for corrective action during processing
    Real time indication of stage of reaction
  • Any combination of the above leads to

Increased profitability

Acknowledgments

  • George A. Fowles -- Retired Vice President of B.F. Goodrich Chemical

Company

- Significant Contribution to Product Concept

  • Karl Oestreich -- B.S. Chemical Engineering

- Significant Contribution on Identifying Applications in the Private Sector

  • Dr. Robert A van Brederode -- Director, Polymer Extension Program,

N.C. State University

- Significant Contribution in Process Implementation

  • Dr. Joseph Bassett -- Retired Professor of Chemistry, Western Carolina University

- Significant Assistance with Experimental Implementation

  • Dr. Roger Bacon -- Dept. Head of Chemistry, Western Carolina University

- Significant Assistance with Experimental Implementation

AccuCure™ Availability

Tison Technologies, LLC has developed a totally automated package, available in a NEMA 4 Stainless steel inclosure using AccuCure™ technology adaptable to different processes.

Exclusive and nonexclusive licenses are available.

CONTACT INFORMATION

AccuCure™ invented by : W. Tison Wyatt, President of Tison Technologies, LLC













AccuCure™ Setup



Gel Checker Setup





Mississippi Polymer Institute, Research Paper

Abstract

This study introduces a new technique to follow the extent of selected polymer reactions using the voltage produced from such reactions. This new measurement process, designated as AccuCure™, uses patented electrodes well shielded from spurious electrical voltages to measure polymerization voltages. In this study, three different types of polymerization reactions were investigated in the evaluation of this new technique. They included one cross-linking free radical reaction, an unsaturated polyester, and two crosslinking polycondensation or step growth reactions that included a polyurethane and an epoxy. The voltages from these reactions were also correlated to the extent of the reactions as indicated from differential scanning calorimeter (DSC) measurements and gel time measurements. This newly discovered electrical phenomenon and the isothermal DSC measurements appeared to closely follow the reaction process for the urethane reaction. In addition, one mirror image AccuCure™ curve was also generated for the urethane reaction consistent with the mechanism proposed for this new electrical phenomenon. The location of the gel time was repeatable and consistent for the urethane reaction relative to the AccuCure measurements. In general, all three polymerization reactions evaluated using the AccuCure™ appeared to have a unique reaction spectrum, and were found to have repeatable and consistent gel time and isothermal DSC measurements. While the cause of this newly discovered electrical phenomenon is not yet fully understood, significant practical commercial implications for monitoring chemical reactions have been demonstrated for the AccuCure™ process.

Introduction

Previous studies performed by Nathaniel Smith and David Shepard (1,2) have followed the cure of thermosetting resins by measuring the dielectric change from an applied voltage (1,2). The dielectric technique is based on the principle that molecular dipoles will align themselves with an applied electrical field. The changes in the measured dielectric properties have been found to be a function of the mobility of the polar groups within a reaction (1,2).

The micro-voltage produced by a chemical reaction is a phenomenon proposed in this study to be a useful analytical tool for observing the progression of a chemical reaction. This technique differs from dielectric analysis in that voltage analysis does not induce a voltage into the reaction medium, it measures the voltage produced from the reaction. The technique uses a patented electrode which is placed into a reaction medium, where the voltage is measured (3).

Tison Wyatt (3) has previously found that as a polymerization reaction proceeds to completion that distinctive changes in potential occur between electrode. In an attempt to increase the understanding of this phenomenon three polymerization reactions were investigated in this study.

1) One cross-linking free radical reaction (Unsaturated Polyester)

2) Two cross-linking polycondensation reaction

One that is moisture sensitive (Polyurethane) One that doesn't generate water (Epoxy)

Recently several articles have appeared in the literature describing isothermal differential scanning calorimeter (DSC) studies of polyurethanes (4), epoxies (5), and polyesters (6). The isothermal DSC study by Rodriguez (6) on unsaturated polyester resins addressed the conversion of double bond functionalities into the single bond matrix of a cross-linked resin. The rate of energy generated by the heat of reaction as measured by an isothermal DSC measurement can then be assumed to be one measure of the rate at which electrons are being transferred during a reaction. Since the new voltage technique is believed to involve the transfer of electrons, then some correlation with the DSC results would be expected. Consequently, isothermal DSC measurements were chosen as a comparison measurement tool for the new voltage evaluation technique.

In this study, polarization reactions involving an unsaturated polyester, a polyurethane and an epoxy were studied for the purpose of correlating the phenomenon of reaction voltage to the reaction's gel time and isothermal DSC curve.

According to Bagotzky and Kisin (7) a battery is an electrochemical device that has two electrodes in contact with an ion-conducting electrolyte. When the electrodes are of two dissimilar materials a potential will exist between the electrodes such that a significant amount of voltage can be produced from the cell. One electrode is termed an anode, the oxidizing electrode where electrons are released and the other electrode is termed a cathode, the reducing electrode where the electrons are absorbed. The anode electrode has a negative polarity and the cathode electrode has a positive polarity. The potential difference between the two electrodes is termed the open circuit voltage (o.c.v.). The o.c.v. is determined by the equation: o.c.v = E(.) - E(-) . Electrochemical principles state that if the two electrodes are of similar materials then the o.c.v. is effectively zero (7). This principle is a key point in the method of measuring the voltage of a reaction. With identical electrodes, one should eliminate the o.c.v. . However, using an AccuCure™ electrode, Tison Wyatt(3) has discovered that a voltage does appear to be produced during many chemical reactions.

Bagotzky and Kisin (7) also indicate that any chemical reaction in which there is a reducer being oxidized, releasing electrons, and an oxidizer that gains electrons has the potential to produce a current. If the oxidizer and reducer are mixed together thoroughly then no electrical energy should be produced. For this case, the electron transfer within the reaction is so randomly spaced that the reaction energy is liberated as heat. A battery, then, produces current because it is able to order the reaction so that a controlled potential is created (7).

Organic Cathodes and Anodes

Cahoon and Heise (8) describe an electrochemical model of organic cathodes and anodes for batteries, in which organic compounds and organic reactions are used to produce electrical current. Cahoon and Fleise (8,9) further classified organic electrodes into the following organic reaction types (9):

1. Oxidations and reductions, including irreversible and redox processes.

2. Substitutions, aliphatic and aromatic, nucleophilic, electrophilic, free radical, etc. in which an atom or group attached to a carbon atom is removed and another enters its place. No change in degree of unsaturation of reactive carbon occurs.

3. Additions of molecules, atoms, or ions to carbon-carbon, and to carbon-hetero multiple bonds to other element unsaturated groups, and ions. These involve an increase in number of groups attached to carbon thus increasing the degree of saturation.

4. Eliminations, which involve a decrease in the number of groups, bound to carbon. The degree of unsaturation increases.

5. Rearrangements, condensations, inversions in which the carbon skeleton of the molecule is internally rearranged.

6. Proton-transfer.

7. Reordering in which one reactant contributes an unshared pair electron to another leaving an empty orbital available. This is simply a Lewis acid-base reaction.

8. Single cleavage.

From the above classifications, one might conclude that the mechanism by which a polymerization would produce a voltage would have many levels of complexity and would vary from one type of polymerization to the next. This would make it very difficult to develop one mechanism that could be used to predict the extent of voltage a reaction would produce for every type of polymerization.

Experimental - Materials

This study utilized three different polymerization reactions including a free radical vinyl ester reaction, an epoxy reaction, and a urethane reaction for the purpose of correlating the phenomenon of reaction voltage to the reaction's gel time and isothermal DSC curve.

These three cross-linking reactions were supplied in a two-part system and mixed together in the required amounts as indicated in their appropriate formulation tables below. Each formulation was formulated to give a room temperature gel time of about eight hours and a gel time of approximately 10 minutes at the elevated isothermal condition of the evaluation in this study.

The unsaturated Polyester reaction involved an oxidative free radical reaction that consists of an unsaturated polyester resin and methyl ethyl ketone peroxide as the initiator. This mixture was evaluated at an isothermal temperature of 65.5 'C or 150 'F. Chemicals are from Bondo Corp., Atlanta, GA (1-404-699-0073).

 

Polyester Resin 98.8%
Methyl Ethyl Ketone Peroxide 1.2%

The Urethane reaction was a polycondensation cross-linking reaction and was moisture sensitive. Chemicals were from Smooth-On Inc. in Gillette, New Jersey 07933 (1-800-762-0744). This mixture was evaluated at an isothermal temperature of 65.5 'C or 150 'F.

 

Smooth-On Crystal Clear 200 Part A 52.63%
Smooth-On Crystal Clear 200 Part B 47.37%

The Epoxy reaction was a polycondensation or stepwise reaction that does not generate a by-product (i.e. water). The epoxy reaction involved is of bisphenol A/epichlorohydrin based epoxy resin cross-linking with a polyamide curing agent (10). This mixture was evaluated at the isothermal temperature of 82.22 'C or 180 'F. Chemicals were supplied from Shell Chemical Company (I -800-TEC-EPON).

 

Eopn Resin 828 52.63%
EPI-Cure 3140 Polyamide Curing Agent 47.37%

Experimental - Method

The electrode used for evaluation was built by Tison Wyatt made from Teflon insulated AccuCure™ electrodes. The probes were submerged into a 10 ml reaction mixture, in a test tube as indicated in Figure 1. These probes were connected to an IBM PC, with a 386 or better CPU, via shielded alligator clips. The computer used utilized a Vernier universal lab interface module capable of measuring +/- 0.1 millivolt and sampled the voltage once per second. Each reaction was run separately to measure the voltage as the reaction progresses. Extra care was required in this study to minimize all possible external sources of voltage from being introduced into the reaction. Shielding was accomplished by removing all electrical sources away from the immediate area of the reaction.

A Sunshine gel checker time meter 22B as shown in Figure 2 was used to determine the gel time of the various cross-linking reactions using the same test tube heating bath as used for the voltage measurements. Each reaction was run separately on the Sunshine gel checker to determine gel time.

The extent of the reaction and the accumulated heat of the reaction were measured versus time using an isothermal Perkin-Elmer DSC-7 measurement method. A 10 mg sample of each reaction was run at the eventual isothermal temperature evaluated on the DSC. The isothermal temperatures of the polymerization reactions were 65.5 'C or 150 'F for the polyurethane and polyester reactions and 82.22 'C or 180 'F for the epoxy reaction. The reaction sample was placed into the DSC at room temperature and raised to 20 'C below the eventual temperature at 40 'C/min. then the DSC acquisition program was started and raised to the isothermal reaction temperature at 40 'C/min. Similar to the procedure used by Rodreiquez (6), the baseline for the cure exothermal was then determined at the isothermal temperature by running the DSC again on the very same sample that was previously used in the DSC isothermal rate.

The voltage from each reaction was then correlated with the extent of the reaction from the DSC and the gel time of the reaction. Each group was repeated three times on each apparatus to determine repeatability.

Results

AccuCure™ and Gel Time Comparison

The AccuCure™ and gel time results for this study are summarized in Figures 3-6 for the polymerizations of the polyurethane, epoxy and polyester respectively. In general, it was found that the results for these three different types of polymerization reactions were very repeatable and consistent.

The results for the AccuCure™ measurements for the three different polyurethane polymerizations in Figure 3 were found to be particularly consistent and reproducible, Note that the second (42) and the third (#3) polymerizations in Figure 3 were found to produce a positive maximum peak at approximately 150 seconds into the reaction at an isothermal temperature of 150 'F. The third polyurethane polymerization (#I) in Figure 3 was found to give a negative peak at approximately 150 seconds into the reaction. This third polyurethane polymerization (#I) was basically a mirror image of the other two positive reaction plots. This result indicates that the polarity of each electrode in a polyurethane reaction was capable of assuming either a positive or a negative charge. All three reactions produced a peak at approximately 150 seconds into the reaction and had a peak width of approximately 400 seconds. Also note that as the AccuCure™ voltage appeared to approach a neutral value asymptotically slightly after 400 seconds. The average of three gel time measurements for the polyurethane reaction was 497 seconds, which is very close to the asymptotic value of time required to achieve the neutral value. Therefore, the gel time appears to be estimable from the time to reach this asymptotic neutral value.

The results in Figure 4 show that the AccuCure™ measurements for the three epoxy polymerizations were again very reproducible and consistent. Mirror image plots, indicative of the passive electrodes switching polarity, were also obtained for the epoxy reaction. The epoxy reaction spectrum as Indicated in Figure 4 appeared to produce at least four reasonably well defined peaks. The first peak was at approximately 75 seconds, the second peak at approximately 375 seconds, the third peak at approximately 500 seconds and the last peak at approximately 600 seconds. The second peak appears to be the predominate peak of the four. The average of three gel times for the epoxy polymerizations was at 563 seconds and was at the approximate location of the fourth peak of the AccuCure™ reaction spectrum.

The first two reactions, polyurethane and epoxy reactions, were condensation or a step wise cross-linking reactions. Odian(10) indicates that these reactions are typically characterized as having monomer or low molecular weight units disappears rapidly at the start of each reaction with each step in the polymerization proceeding by approximately the same rate. Thus as reactive sites are used up the polymerization rate slows down. This suggests that as the reaction mechanism proceeds that the reaction becomes ordered quickly and is able to develop anodic and cathodic domains that are detectable by AccuCure™ as voltage.

Figures 5 and 6 show the results of the AccuCure™ measurements for the polyester polymerization. The polyester polymerizations were somewhat less consistent than the previous polyurethane and epoxy reactions. There appears to be a negative peak in Figure 5 at the start of the plot around 50 seconds, but the rest of the results seem to be less predictable in the short time frame of this evaluation. To check this hypothesis an extended polyester polyrneri2,ation was generated and the results from this run are shown in Figure 6. The average of three get times for the polyester reactions was 597 seconds. As indicated in Figure 6 the average gel time and the peak of the extended run were apparently near the same time frame. This suggests that the evaluation of the polyester polymerization using the AccuCure™ process may be more effective if the results for this material were taken over a longer period of time to follow the effectiveness of the reaction process.

This last polyester reaction was also a free radical cross-linking reaction, Salla and Ramis(11) indicate that this reaction is typically characterized as having a very complex reaction process with many reactive processes occurring simultaneously (i.e. Initiation, propagation, and termination). At any instance the reaction mixture could contain monomer, high molecular weight polymer or growing chains. Given this reaction mechanism one might conclude that a free radical cross-linking reaction might not order as quickly as a step polymerization, thus requiring a longer time frame for consistent voltage measurements.

The plots for the polyurethane, epoxy, and polyester reactions appear to be unique for each reaction. This could suggest that the AccuCure™ technique might be a useful spectrum analysis tool for the study of or identification of various types of reaction processes.

DSC Measurement Compared to AccuCure™ Electrode Measurement

As discussed in the introduction, the rate of energy generated by the heat of reaction as measured by an isothermal DSC measurement can be assumed to be one measure of the rate at which electrons are being transferred during a reaction. Since the AccuCure™ process is believed to involve the transfer of electrons, then some correlation with the DSC results would be expected. Consequently, isothermal DSC measurements were chosen as a comparison measurement tool for the new voltage evaluation technique. The results summarizing the comparison between the DSC measurements and the AccuCure™ electrode measurements are summarized in Figures 7-9.

The isothermal DSC curve for the polyurethane reactions as indicated in Figure 7 indicates that the maximum DSC peak and the maximum for the AccuCure™ peak occur at approximately the same time. With compensation for differences in sample heat up rates to isothermal conditions and different sample sizes, timing to the peak maximum could be considered identical for both measurements. This result suggests that both techniques are measuring the same polyurethane reaction characteristic.

The AccuCure™ curve for the epoxy reaction in Figure 8 suggests many different reaction stages, where as the DSC curve only indicates one reaction peak. However, the first peak of the AccuCure™ curve does appear to correspond with the maximum in the DSC curve. This suggests at least one epoxy reaction component appears to be similarly indicated by both the DSC and the AccuCure™ results.

Finally, Figure 9 compares an isothermal DSC curve with the AccuCure™ curve for one of the polyester reactions evaluated. Apparently the AccuCure™ and DSC techniques are measuring different stages of the reaction, since the maximums are occurring at different locations. This result suggests that the DSC and the AccuCure™ results do not indicate a common reaction process for the polyester reaction.

Discussion of voltage Phenomenon in Polymer Reactions

It is proposed that the following conditions would be expected to exist in-order for voltage to be produced and measured from a polymerization reaction with similar electrodes:

1. Anodic like and cathodic like domains should exist within the reaction, (a domain is considered to be a specific volume of the reaction mixture)

2. The domains would be significant enough in size to produce a measurable voltage.

3. The domains would be ordered such that a net potential difference is produced (i.e. a non-random system).

4. It appears that if the anodic and cathodic domains are too random and/or insignificant in size that a voltage may not be detectable.

5. Thermodynamic conditions should be conducive to the production of electrical energy.

6. The reaction and electrodes must be shielded from spurious electrical voltage from equipment in the vicinity.

A possible explanation of the observed voltages produced with AccuCure™ electrodes is as follows. Polymerization starts out in a manner where the reactants are possibly randomly ordered so that the reaction does not produce an initial voltage and as the reaction proceeds, if the reactants stay randomly oriented, the reaction would only produce heat. However, as the reaction proceeds and produces a polymer of significant mass, the reaction may become ordered such that a voltage is produced. At this point the reaction medium consists of a polymer mass, monomer, and other reactants. It is proposed that the polymer mass possibly during the reaction becomes ordered in such a way that a potential difference is created and measurable by a voltmeter. As the polymerization proceeds to completion the potential difference should disappear. The above is proposed to explain how the phenomenon of voltage produced by a reaction might occur. The critical point here is even if we do not understand this phenomenon, the measurable change in voltage during a reaction should have significant application if it relates to a reaction consistently and repeatably. It will be left to future researchers to determine the mechanisms responsible for this phenomenon.

Conclusion

This study utilized three different polymerization reactions included a free radical vinyl ester reaction, an epoxy reaction, and a urethane reaction for the purpose of correlating the phenomenon of reaction voltage to the reaction's gel time and isothermal DSC curve. A new technique was introduced to follow the extent of selected polymer reactions using the voltage produced from such reactions. This new measurement process, designated as AccuCure™ uses patented electrodes well shielded from spurious electrical voltages to measure polymerization voltages.

In general, all three polymerization reactions evaluated using the AccuCure™ appeared to have a unique reaction spectrum, and were found to have repeatable and consistent get time and isothermal DSC measurements.

This newly discovered electrical phenomenon and the isothermal DSC measurements appeared to closely follow the reaction process for the urethane reaction. In addition, one mirror image AccuCure™ curve was also generated for the urethane reaction. This result indicates that the polarity of each electrode in a polyurethane reaction was capable of assuming either a positive or a negative charge. The average gel time for the polyurethane reaction was very close to the asymptotic value of time required to achieve the neutral value. Therefore, the gel time can be estimated from the time to reach this asymptotic neutral value. The isothermal DSC curve for the polyurethane reactions indicates that the maximum DSC peak and the maximum for the AccuCure™ peak occur approximately the same time. With compensation for differences in sample heat up rates to isothermal conditions and different sample sizes, timing to the peak maximum could be considered identical for both measurements. This result suggests that both techniques are measuring the same polyurethane reaction characteristic.

The results for the epoxy polymerizations were again very reproducible and consistent. Mirror image plots, indicative of the passive electrodes switching polarity, were also obtained for the epoxy reaction. The epoxy reaction spectrum appeared to produce at least four reasonably well defined peaks. The average gel time for the epoxy polymerizations was approximately identical with the fourth peak of the AccuCure™ reaction spectrum. The AccuCure™ curves for the epoxy suggests many different reaction stages, where as the DSC curve only indicates one reaction peak. However, the first peak of the AccuCure™ curve does appear to corresponds with the maximum in the DSC curve. This suggests that at least one epoxy reaction component appears to be similarly indicated by both the DSC and the AccuCure™ results.

The polyurethane and epoxy reactions, are condensation or step wise cross-linking reactions which suggests that as the reaction mechanism proceeds that the reaction becomes ordered quickly and is able to develop anodic an cathodic domains that are detectable by AccuCure™ as voltage.

The polyester polymerizations were somewhat less predictable in the short time frame used for the other polymerization reactions. To check this hypothesis all extended polyester polymerization was generated and the average gel time and the peak of the extended run were near the same time frame. This suggests that the evaluation of the polyester polymerization using the AccuCure™ process may be more effective if the results for this material were taken over a longer period of time to follow the effectiveness of the reaction process. The results for the polyester reactions indicate that the AccuCure™ and DSC techniques are measuring different stages for the polyester reaction since the maximums occurred at different locations. This result suggests that the DSC and the AccuCure™ results do not indicate a common reaction process for the polyester reaction.

This last polyester reaction was also a free radical cross-linking reaction indicating that this reaction is typically characterized as having a very complex reaction process with many reactive processes occurring simultaneously. Given this reaction mechanism one might conclude that a free radical cross-linking reaction might not order as quickly as a step polymerization, thus requiring a longer time frame for a consistent voltage measurement.

The AccuCure™ process results obtained in this study suggest that this process involves the detection of the transfer of electrons during a polymerization reaction. In this process polymerizations, in general, start out in a manner where the reactants are possibly randomly ordered so that the reaction does not produce a voltage. As the reaction proceeds, if the reactants stay randomly oriented, the reaction would only produce heat. However, as the reaction proceeds it is proposed that the polymer mass possibly orders the reaction such that a potential difference is created and measurable by a voltmeter. As the polymerization proceeds to completion the potential difference should disappear. In general, the plots for the polyurethane, epoxy, and polyester reactions appear to be unique for each reaction. This would suggest that the AccuCure™ technique might be a useful spectrum analysis tool for the study of or identification of various types of reaction processes.

While the cause of this newly discovered electrical phenomenon is not yet fully understood, significant practical commercial implications for monitoring chemical reactions have been demonstrated for the AccuCure™ process. It will be left to future researchers to determine the mechanisms responsible for this phenomenon.

Reference

I . Nathaniel Smith, and David Shepard, "Dielectric Cure Analysis: Theory and Industrial Applications", Reprint from SENSORS, October 1995.

2. John Lane, and James Seferis, "Dielectric Studies of the Cure of Epoxy Matrix Systems", Journal of Applied Polymer Science, Vol. 31, 1155-1167, 1986.

3. Tison Wyatt, Patent Owner.

4. Pelguang Zhou and H.L. Frisch, ''Isothermal Reaction Kinetics and Phase Behavior Analysis In the Formation of PCU/PMMA interpenetrating Polymer Networks", Macromolecucles 27, 1788-1794. 1994.

5 . E.M. Woo and J.C. Sereris,"Cure Kinetics of Epoxy/Anhydride Thermosetting Matrix Systems", The Journal of Applied Polymer Science, Vol. 40, 1237-1256. 1990.

6. Ernesto L. Rodriguez, "The Effect of Free Radical Initiators and Fillers on the Cure of Unsaturated Polyester Resins", Polymer Engineering and Science, vol. 31, No. 14. July 1991.

7. V.S. Bagotzky and A.M. Kisin, Chemical Power Sources, Academic Press, New York, 1980.

8. N.C. Cahoon and G.W. Heise, The Primary Battery, John Wiley and Sons, New York, 187 -238.1976.

9. N.C. Cahoon and G.W. Heise, The Primary Battery, John Wiley and Sons, New York, 201. 1976.

10. G. Odian, Principles of Polymerization,, John Wiley and Sons, New York. 1981.

11. J.M. Salla and X. Ramis, "A Kinetic Study of the Effect of Three Catalytic Systems on the Curing of an Unsaturated Polyester Resin", Journal of Applied Polymer Science, Vol. 51, 453-462. 1994.


Western Carolina University
Research Report
Avery Research Report
Avery Graphs



Background:

In the past eighteen months, I have conducted a series of tests utilizing stainless steel and graphite electrodes built by Tison Wyatt to monitor a voltage that appears to be produced by the progression of an aqueous reaction. These tests were conducted at Western Carolina University under the supervision of Dr. J. Roger Bacon, professor of chemistry and Department Head. An experimental electrode system developed by Mr. Tison Wyatt was employed to monitor the observed voltages.

Apparatus:

The initial experimental apparatus consisted of a special stainless steel electrode connected to a MPLI .(Vernier Software, Portland, OR) computer interfacing board installed in an IBM compatible computer for collection of the experimental data. The software supplied with the card was used to set the scaling and timing for the data collection. A 10K resistor was connected between the electrode leads and the interface box to reduce the noise that accompanies analog signals such as these.

The test reactions were carried out in Pyrex beakers that along with the electrodes, were cleaned using methanol followed by three rinsings with distilled water. Mr. Wyatt provided the electrodes used for these tests. The stainless steel electrodes were 2.4 mm in diameter. A sealed Teflon coating was permanently placed around each electrode 26 mm above the end of the stainless steel core. This was done so equivalent electrode surface areas could be maintained.

After obtaining some of the initial data with the stainless steel electrodes, graphite electrodes were used to rule out the possibility that the observed voltages could only be detected using metal electrodes. The graphite electrodes were also coated and sealed using Teflon.

Experimental:

The reaction used to test this new electrode system was the aqueous oxidation reduction reaction between sodium bisulfite (NaHSO3) and potassium iodate (KIO3).

IO3-1 + 3 SO3-2 ------------> I- + 3 SO4-2

IO3-1 + 5 I- + 6 H+ ------------> 3 I2 + 3 H2O

As shown, iodine is one of the reaction products. A well known characteristic of iodine is its formation of intensely colored compounds. The reaction shown above begins with two colorless solutions and upon completion yields an orange solution. The production of a colored product is beneficial in that it allows visual monitoring of the reaction progression in conjunction with the voltage data obtained by Mr. Wyatt's electrode system.

These tests were conducted using 60 ml. of 0.1 M. KIO3 and 15 ml. of 0.025M. NaHSO3. The reactions were carried out at room temperature. Also, the combination sequence of the reactants was always the addition of the NaHSO3 to the KIO3. The purpose of doing so was to facilitate the mixing of the reactants since mechanical stirring was not utilized. The reactant stock solutions were prepared freshly each day, and were kept in sealed flasks. Serial dilutions were used to obtain the desired concentrations.

Results:

The first set of tests were conducted using stainless steel electrodes attached to the Vernier MPLI. The general trend for this set of tests was as follows. After addition of 60 ml of 0.1 M. KIO3 to 15 ml. of 0.025M. NaHSO3, an initial baseline potential of approximately 0.009 volts was established and then began to decrease slowly. At 71 seconds it reached its minimum potential of 0.0065 V. The solution began to change from colorless to orange. (This color change was expected due to the production of dissolved iodine, as referred to previously.) The potential rapidly increased and reached a maximum of 0.011 V. around 78 seconds. The potential of the solution then decreased, returning to a value near that of the system prior to the onset of the chemical reaction by 130 seconds.

Analogous tests were conducted using the graphite electrodes. The change of color was again observed around 71 seconds. However, there was a significant difference in the behavior of the baseline voltages prior to reaction's start. Instead of a distinct decrease in the potential occurring, the baseline voltage remained almost constant prior to the onset of the redox reaction. From a voltage of 0.0065 V. at 71 seconds, the potential of the aqueous solution rapidly increased to a maximum of 0.009 V. at 91 seconds. The solution potential remained at this potential value for the remainder of the testing cycle.

Each system was tested repeatedly within the given set of conditions. A level of note worthy reproduction was obtained for each of the electrodes. Furthermore, the initial significant increase of potential for these aqueous systems, always preceded the first indication of a color change, or visual verification that the redox reaction was occurring.

Conclusions:

1. A small change in voltage was detectable as a result of the oxidation reduction reaction between potassium iodate and sodium bisulfite.

2. The time of and magnitude of the observed voltage are reproducible for given reactant solution concentrations and set conditions.

3. The voltage produced by the aqueous reaction can be followed using stainless steel or graphite probes made by Tison Wyatt. This was unexpected since graphite (carbon) is considered a very poor conductor or electricity.

4. The ability to monitor the voltage produced by these aqueous reactions utilizing graphite strongly opposes the assumption that the observed phenomena is due to metallic interactions introduced by the utilization of the stainless steel electrodes.

5. It appears that it is possible to monitor the progression of an aqueous chemical reaction, as well as determine its starting and completion points, utilizing Mr. Tison Wyatt's electrode system.


Avery D. Fox











DEPARTMENT OF CHEMISTRY AND PHYSICS

(704) 227-7260

January 16, 1997

To whom it may concern:

I recently supervised the work of a student, Mr. Avery Fox, who ran a series of test on a new electrode system for determining the progress of chemical reactions. The work was done to demonstrate the process developed by Mr. Tison Wyatt. Mr. Wyatt's brother Tracy Wyatt contacted us to have some tests run. We agreed to carry out the work here at Western Carolina University.

Mr. Avery Fox is a dean's fist student here at WCU and is a second semester junior. He has excellent laboratory skills and is a diligent worker. I have observed his work, discussed his experiments, and helped him evaluate the results obtained. I have confidence in the work he has done on this project.

The chemical reaction used to study the electrode response is the oxidation-reduction reaction between potassium iodate (KI03) and sodium bisulfite (NaHS03). The reaction is slow enough so timing the reaction time is not complicated. The reaction also has dissolved iodine I2 as one of the final products. This allows us a visual indication of when the reaction has reached completion. The two reaction are indicated below. The first is slow and goes to completion before the fast second reaction occurs.

IO3-1 + 3 SO3-2 ----> I- + 3 SO4 -2

IO3-1 + 5 I- + 6 H+ ----> 3 I2 + 3 H2O

The electrodes system was connected to a MPLI (Vernier Software, Portland, OR) computer interfacing card installed in an IBM compatible computer for collection of the experimental date. The software supplied with the card was used to set the scaling and timing for the data collection.

Sincerely,

J. Roger Bacon

Professor and Department Head


CULLOWHEE, NORTH CAROLINA 28723-9050

Western Carolina University is one of the sixteen senior

institutions of The University of North Carolina and an

affirmative action/equal opportunity employer.

Dynamic Systems, Inc.
Statement
Graphs





Dynamic Systems, Inc.
Development Facility
46 Sunlight Dr.
Leicester, NC 28748


704-693-3523


 

Tracy Wyatt
Tison Technologies, LLC
Box 1268
Skyland, NC 28776

Dear Mr. Wyatt,

Since November 1996, we have been running experiments with the Tison Technologies, LLC

AccuCure™ voltage sensing device to monitor polyurethane foaming reactions. Voltage vs. time has been graphically monitored and a voltage-time output has been observed and recorded that seems to show a repeatable result with change in the formulation. We are continuing to examine this method of monitoring chemical reaction since we think it may have some possibilities. There are also a number of variables which still need to be resolved. The phenomenon is real, however additional tests are required in order to determine the reason and mechanism for the voltage variations. We consider this technique worth careful study.

Sincerely,


Charles A. Yost
President





AccuCure™ Research and Reporting


To review some of the research done on the AccuCure™ system developed by Tison Technologies, LLC, choose the research facility you wish to review from the menu above. We are continually having our products tested and more reseach will be posted in the future.

Copyright © 2003 Tison Technologies, LLC. All Rights Reserved.